As the principal hamodynamic parameters mentioned before form an integral and essential part of the evaluation of heart disease, Physio FlowTM can be used in several clinical conditions :
Acute
and chronic heart failure and their evolution (low stroke volume and
ejection fraction together with elevated preload are markers of unfavorable prognostic)
Promotion
of optimum drug therapy
Critical
care


cyclo-ergometer
or treadmill exercise testing (cardiology and sports medicine)
Reduced
baroreflex sensitivity, which is a marker of unfavorable autonomic balance
(PhysioFlowTM gives a non-invasive estimate of reactivity and activity of
the autonomic nervous system when using cardiovascular reflex testing such as
tilt-test, Valsalva maneuver, or orthostatic tests like postural changes).
Hypertension
Dialysis
monitoring
Optimal
adjustment of A/V and V/V delays in cardiac resynchronization therapy
Research
(pharmacology, physiology, aerospace)
Etc.
Even in the case of any trouble of cardiac rhythm, accurate results are provided without being influenced by the type of trouble.
PhysioFlowTM is also designed for research work and can be applied to any population or group of people presenting risk of heart diseases whatever the age or morphology.
Clinical studies in PDF format are available in the download section of this website.
As a cardiac output measuring system, PhysioFlowTM is validated against generally accepted reference methods, such as "direct" Fick, thermodilution and Echo-Doppler. Validations have been obtained at rest and during exercise on cyclo-ergometer, on healthy subjects and patients suffering from cardiac or pulmonary diseases. Reproducibility has also been tested thoroughly.

1) Reproducibility
1st
test
hemodialysis patients, 3 sessions, (75 measurements on 25 patients)
measurements performed before ultrafiltration
measured parameter : stroke volume that is less sensitive to heart rate variations than cardiac output
Results : excellent reproducibility (r = 0.98 p < 0.0001)
2nd
test
Similar results were obtained during another test :
7 patients at rest (lying), 3 operators, 3 consecutive measurements made by each operator (device completely switched off between measurements, different electrode brands used). The purpose was to evaluate reproducibility of stroke volume measurements under various measurement conditions and operators.
Results : excellent inter- intra-operator reproducibility (r > 0.96, p<0.0001)
It has also been shown that reproducibility is remarkable during cyclo-ergometer exercise, and under non-optimal measurement conditions (important modifications of electrode positions).
2) Validation against "direct" Fick method (Eur 5 Appl Physiol(2000) 82: 313-320)
Exercise by a new impedance cardiography device: comparison with the "direct" Fick method
Anne CHARLOUX, MD, Eve LONSDORFER-WOLF, MD, Eliane LAMPERT, MD, PhD, Monique OSWALD-MAMMOSSER, MD, Bertrand METTAUER, MD, PhD, Bernard GENY, MD, PhD, and Jean LONSDORFER, MD. Service des Explorations Fonctionnelles Respiratoires, Cardio-Circulatoires et de l'Exercice, Hôpitaux Universitaires de Strasbourg, Hôpital Civil, BP 426, 67091 Strasbourg Cedex, France.
Study objectives : To evaluate the reliability and accuracy of a new impedance cardiography device, the PhysioFlowTM, at rest and during a mild constant effort performed in the supine position.
Design : We compared cardiac output determined simultaneously by two methods, the PhysioFlowTM and the "direct" Fick methods.
Patients : Forty patients referred for right cardiac catheterization, 14 with sleep apnoea syndrome and 26 with chronic obstructive pulmonary disease. All patients had normal left ventricular function.
Measurements and results : Thirty two patients have been able to perform an exercise, ranging from 10 to 50 watts, 53% of patients pedaling at a 30 W workrate. The 72 cardiac outputs measured at rest and during exercise range from 4.34 to 14.84 L/min for Physio Flow and from 3.60 to 15.03 L/min for "direct" Fick. VO2 range from 0.14 to 1.19 L/min. Regression of QcPF (L/min) against VO2 (L/min) is shown in fig 2. The regression equation, based on 72 measurements, is QcPF = 7.2 O2 + 4.2. Individual values of cardiac output obtained concomitantly by the two methods (QcFick and QcPF) are plotted in fig 1. The correlation coefficient was 0.89 (p<0.001) for resting values and 0.85 (p<0.001) for exercising values. There was no difference between cardiac output means measured by the two methods, either at rest (mean QcPF = 6.25 ±± 1.14 L/min, mean QcFICK = 6.32 ±± 1.42 L/min, NS) or during exercise (mean QcPF = 9.89 ±± 2.17 L/min, mean QcFICK = 10.15 ±± 2.35 L/min, NS).
Comparison of the two techniques with the Bland and Altmann method is represented in fig 2. At rest, the mean difference (QcFICK-QcPF) was 0.07 L/min (95% confidence interval: -0.23, +0.27 L/min), which represents 0.29% of the mean resting cardiac output. The (QcFICK-QcPF) difference did not vary significantly with the mean of (QcFICK, QcPF). The limits of agreement are (-1.18, +1.32 L/min) at rest. Thus, in 95% of paired measurements, the difference between QcFICK and QcPF is within a (-19%, +19%) interval. During a mild effort, the mean of the (QcFICK-QcPF) difference is 0.26 L/min (2.44% of mean exercising cardiac output) with a (-0.17 , +0.69 L/min) 95% confidence interval. The limits of agreement during exercise are -2.16 and +2.68 L/min (-21%, +26% of cardiac output). In all patients, the direction of the changes in impedance cardiography values always corresponded to those measured with "direct" Fick. At rest, QcFICK and QcPF differed by 20% or more in only one patient among the 40 (2.5%). During exercise, two (QcFICK-QcPF) differences among the 32 measurements are equal or are greater than 20% (6.2%).
Conclusions : Physio Flow provides a clinically acceptable evaluation of cardiac output in these conditions. This new impedancemetry device deserves further study in other populations and situations.
Fig1 : Correlation graphs : At rest (n = 40)

Fig2 : Bland and Altmann scattergrams : Exercise (n = 32)

3) Validation against Echo-Doppler
Use of new thoracic impedance method to assess cardiac index in children
(Published at the American Society of Anesthesiology, Scientific Sessions, Orlando 1998)
E Wodey, MD, X Beneux, MD, F Carre, MD, O Azzis, MD, C Ecoffey, MD. Depts of Anesthesiology 2 and Physiology , Hôpital Pontchaillou, 35033 Rennes, France.
Introduction: Measurements of cardiac output give important information about hemodynamic status, and it can be evaluated by several methods. In 1966 Kubicek et al described a method of calculating stroke volume, i.e. cardiac output, from the thoracic impedance signal. However, special electrodes, specific configuration, skin cleansing, and measurements for estimation of one cylinder or truncated cone model, were essential to obtain the true value of stroke volume, but they could also induce some error in this assessment. Recently a new thoracic impedance method Physio Flow® (Manatec Biomedical, Macheren, Paris) has been available. Unlike traditional impedance methods, positioning of electrodes is not critical, nor are special eletrodes or skin cleansing. In addition, no estimation of cylinder or truncated cone model is needed. Thus, the aim of this study was to evaluate the ability of Physio Flow® to determine cardiac index in healthy children.
Methods: Cardiac index measurements were performed in 24 healthy children (median age 11, range 6-16 yrs) in a double-blind manner by echocardiography-Doppler Sonos 1000™ (Hewlett-Packard, Andover), and by Physio Flow™ in a supine position at rest. In all cases informed parental consent was obtained. Aortic annular diameter (Øao) and flow velocity in the ascending aorta were measured. Cardiac Index (CIdop) was calculated from the volumetric equation: CIdop (L.min-1.m-2) = Vao (cm.sec-1) x (pix(FIao x 0.5)2)(cm2) x 0.06 / BSA (m2). Concerning the new bioimpedance thoracic method, six standard electrodes were applied on children without specific skin preparation, provided that one pair was at the thorax base and another was at the thorax apex. Measured parameter were maximum rate of change of transthoracic impedance (dZ/dtmax) and the thoracic fluid inversion time (TFIT). According to the manufacturer, there was a linear relationship between the following combined parameters: stroke volume/(BSA x pulse pressure) and TFIT/heart rate, allowing estimation of the calibration stroke volume. After calibration, stroke volume varied according to the variations of TFIT and dZ/dtmax. Spearman's rank test was used for the correlation studies.
Results : Cardiac index quantified by continuous wave Doppler (CIdop) was significantly linked to cardiac index quantified by PhysioFlowTM(CIimp) (Figure A). An underestimation of cardiac index was obtained by the PhysioFlowTM method for all measurements. The overall bias between CIdop and CIimp was +1.28 L.min-1.m-2. The 95% limits of agreement between both methods was ±0.96 L.min-1.m-2 (Figure B).
Discussion: CI may be measured with reliability from PhysioFlowTM in healthy children older than 6 years of age at rest, although systematic under-estimation was found. Reference: Kubicek WG et al :Aerospace Med 1966 ;37 :1208-1212

4) Detection of coronary artery disease during cyclo-ergometer exercise, using new generation impedance cardiography (*)
(*) This abstract displays results that will be thoroughly analyzed and published later
Objectives:
The protocol was designed to determine if a new non-invasive cardiac output measuring device, called PhysioFlowTM, based on analysis of instantaneous thoracic impedance variations (new generation impedance cardiography ICG), could be helpful to detect Coronary Artery Disease (CAD) during cyclo-ergometer exercise.
We have used the following criteria for positivity of ICG during cyclo-ergometer exercise (figure 1) : stroke volume profile (SV) alteration : SV decreases at a certain level of effort after having increased, whereas heart rate continues to increase. The result is that at the end of exercise, SV is not maximal.
Methods:
1)
Patients
Heart rate and stroke volume determinations were performed during cyclo-ergometer exercise on a group of 31 patients, as well as simultaneous exercise ECG. These patients were referred to our unit because of various clinical symptoms (chest pain, dyspnea). All patients were measured whatever their morphology, age, or sex . All patients who represented positive cyclo-ergometer exercise or clinical symptoms (chest pain and / or dyspnea) have undergone control coronary angiography (19 men, 12 women, age = 58 years ±10 years).
2)
Materials
Exercise ECG : Quinton 5000 series.
The ICG method (PhysioFlowTM ) is a new impedance method based on analysis of instant thoracic impedance variations, using six electrodes (two for ECG (measurement of heart rate) and four for ICG). Measured parameters are heart rate (HR), contractility index (CI), and Thoracic Flow Inversion Time (TFIT) (active period of left ventricular ejection). (SV) is derived from CI and TFIT. Cardiac Output (CO) = HR x SV.
3)
Procedures
Patients underwent a classical cyclo-ergometer exercise in the same hospital cardiology unit, with steps of 30 watts every 2 minutes. ICG was calibrated just before exercise started, and recorded measurements every 12 heart beats (mean value of 12 previous beats). Exercise was stopped when patients displayed clinical signs of CAD, pain, dyspnea, or ECG ST-segment depression.
We performed control coronary angiography, as it is considered the gold standard for this study.
Results:

Discussion:
Only one patient had positive control coronary angiography and negative ICG and ECG examination, because he had only 40% stenosis of the right coronary artery and was receiving betablocker treatment, making it difficult to detect CAD during exercise. In the remaining cases (30), (where EKG, ICG, and control coronary angiography were all positive), stroke volume profile alteration always occurred before EKG ST-segment depression.
Overall, EKG was not consistent with control coronary angiography 8 times out of 31, and ICG only one time in 31. In 2 more cases the EKG could not be interpreted because of conduction defects (ICG and control coronary angiography were both positive in these cases).
Conclusion:
Considering these results, stroke volume profile measurements performed by ICG seem to be a useful tool in association with EKG for detecting CAD during cyclo-ergometer exercises. They help making a decision concerning control coronary angiography prescription, especially when EKG gives a negative or unclear conclusion. It is also useful to assess hemodynamic impairments when related to CAD and in the setting in of heart failure or other diseases like Syndrome X (when simple control coronary angiography cannot show micro circulation abnormality).
Figure 1 : Normal and abnormal stroke volume profile (before/after dilatation)

5) Other results :
Reference method : Echography - Doppler
measurements performed only at rest
19 patients (4 normal control subjects, 15 patients suffering from ischemic disease or cardiac insufficiency)
measured parameter : cardiac output
n = 19 ; R = 0.83 ; p < 0.001
Reference method
: VO2 and CO2 rebreathing
8 healthy subjects : exercise up to 180 watts, steps : 30 watts
measured parameter : cardiac output by Physio Flow versus CO2 rebreathing and VO2
n = 48 ; R = 0.85 ; p < 0.001
A new concept for diagnosis and therapy follow up.
RSVI : systemic vascular resistance index (dyn.s.cm/m²)
LCWI : left cardiac work index (kg.m/m²)
A graphical representation of vascular resistance / cardiac work equilibrium
A diagnosis tool (classification and quantification of physio-pathological status), making also drug therapy follow up easier.
Case n°1 : Cardiac
insufficiency
Results (after four days of treatment) :
Significant improvement of stroke volume ( + 40 % ) as well as cardiac output leading to a left cardiac work index increase.
A decrease of vascular resistance index due to the reduction in blood pressure.
Case n° 2
: Hypertension
Marked reduction of blood pressure and peripheral vascular resistance index as well
No increase of left cardiac work index ; consequently no change in myocardial oxygen consumption
Case n°3 :
Detection of coronary diseases during cyclo-ergometer exercises
no improvement of stroke volume at rest
significant improvement of exercise stroke volume profile associated with disappearance of angina and dyspnea

Case n° 4
: Optimal dual chamber pacemaker AV-delay adjustment
Best stoke volume is obtained with a 140 ms AV-Delay
Difference between best and worst stroke volume is 12 ml (23 %)
Case n° 5
: Cardiac insufficiency, effect of drug therapy on exercise performance
Context :
Two consecutive exercise-tests on cyclo-ergometer (30W, 3 min, measurements obtained by mean values of 12 consecutive heart beats), patient suffering from cardiac insufficiency resulting from myocardial infarction.
Patient benefited from digitaline and nitrate drug therapy between two exercises (1 month between each exercise test).
Results :
First exercise : low stroke volume at rest, stroke volume decrease during exercise, early end of exercise due to dyspnea .
Second exercise : better stroke volume at rest (+ 23 %), better stroke volume profile during exercise, exercise lasted longer, dyspnea appeared later.
Case n° 6
: Critical care monitoring, effect of dobutamine and noradrenaline
Context :
Critical care unit of a University Hospital
two patients with a different hemodynamic status, the first needing a dobutamine perfusion, the second noradrenaline
Results :
Patient 1 displays a recovery of his cardiac index under dobutamine (from 2.0 l/min/m² to 3.7 l/min/m²)
Patient 2 displays a decrease of cardiac index under noradrenaline (from 5.5 l/min/m² to 2.9 l/min/m²)
Case
n° 7 : Critical care diagnosis ; septic shock and cardiac shock
Context :
Emergency unit of a general hospital, at night
Two emergency patients, no patient file or data
Clinical assessment is unclear, very low blood pressure (about 80/55 mm Hg) and high heart rate (about 120 bpm)
Need for an urgent hemodynamic status assessment
Results :
Patient 1 displays a high cardiac output/index, associated with very low vascular resistance index and more or less normal left cardiac work index : diagnosis was septic shock.
Patient 2 displays a low cardiac output/index, associated with very low stroke volume, low vascular resistance index, and low left cardiac work index : diagnosis was in favor of a cardiogenic shock.
Physioflow : a new era in continuous noninvasive cardiac output monitoring. Physioflow performs continuous cardiac output measurements in various clinical conditions. Physioflow is an impedance cardiograph featuring enhanced bioimpedance signal morphology analysis.